Monte Carlo study of the scattering error of a quartz reflective absorption tube
Piskozub, J., P.J. Flatau, and J.V.R. Zaneveld. Monte Carlo study of the scattering error of a quartz reflective absorption tube. Journal of Atmospheric and Oceanic Technology, V18(N3):438-445, 2001.
Read / download PDF (0.2 MB)Full article text
Text extracted from the original PDF. Read the PDF for figures, equations and original formatting.
PDF page 1
arXiv:physics/0006061v1 [physics.ao-ph] 23 Jun 2000
Monte Carlo study of the scattering error of a quartz
reflective absorption tube
Jacek Piskozub
Institute of Oceanology PAS
Powsta´nc´ow Warszawy 55 81-712 Sopot, Poland
piskozub@iopan.gda.pl
Piotr J. Flatau
Scripps Institution of Oceanography, University of California, San Diego
La Jolla, CA 92130-0221
pflatau@ucsd.edu
J. V . Ronald Zaneveld
College of Oceanic and Atmospheric Sciences, Oregon State University
104 Ocean Admin Bldg Corvallis, OR 97331-5503
October 30, 2018
In Press: Journal of Atmospheric and Oceanic Technology, 2000
Abstract
A Monte Carlo model was used to study the scattering error of an absorption meter with a
divergent light beam and a limited acceptance angle of the re ceiver. Reflections at both ends
of the tube were taken into account. Calculations of the effe ct of varying optical properties
of water, as well as the receiver geometry, were performed. A weighting function showing
the scattering error quantitatively as a function of angle w as introduced. Some cases of the
practical interests are discussed
1 INTRODUCTION
Light absorption is the essential element required for mari ne phytoplankton growth. The re-
gion of the upper ocean illuminated by sunlight is responsib le for most of marine primary
production. Absorption and scattering of phytoplankton co ntrol the light field (at least for
1
PDF page 2
Case 1 waters which comprise a large majority of ocean waters ), and thus, affect the spectral
reflectance of the ocean surface. Such spectral changes provide information about ocean color
(Gordon and Morel, 1983)
In all radiative transfer modeling of the marine environment the inherent optical properties
(IOPs) are either the necessary input parameters, or in the case of inverse problems, the output
of the calculations. This makes them important in marine opt ics studies. The measurement of
the attenuation coefficient is relatively easy; an attenuation meter consists of a collimated light
source and a collimated light detector at a known distance. T he only major problem in the
attenuation measurement are photons from outside sources or the ones reflected on the instru-
ment scattered into the light beam, causing an underestimat ion of the attenuation coefficient
value. The detector has a finite aperture so that some of the fo rward scattered photons will
be counted as part of the direct beam, also causing an underes timation. However, the mea-
surement of the components of the attenuation, the absorpti on and scattering coefficients, is
inherently much more difficult, and researchers have been st riving to minimize the measure-
ment errors of these parameters for most of this century.
The central idea behind any absorption measurement is to pro ject a beam of light through
an absorbing medium. If one could measure all of the unabsorb ed light of the direct beam,
as well as all of the scattered light, the only light lost woul d be the absorbed light. Thus,
absorption meters tend to be arranged to collect as much of the scattered light as possible. The
absorption coefficient measured in any optical device has tw o major sources of error, both of
which are due to the fact that natural suspensions tend to sca tter light as well as absorb it.
In practice, the scattered light traverses a longer path thr ough the absorbing medium and so
is more likely to be absorbed (the path length amplification f actor). Secondly, not all of the
scattered light is collected due to the geometry of the absorption meter (the scattering error).
One of the choices to measure in situ absorption is the cylind rical reflection tube. Such
a tube needs to be be long enough to provide a sufficient optica l path to measure the low
absorption values typical for Case1 waters below 580nm. Any photon scattered forward off
the instrument axis is, at least in theory, reflected by the tube walls until it reaches the detector.
An ideal instrument of this kind should also have a reflector inside of the source end to collect
the backscattered photons.
The first working prototypes of such a device were developed by Zaneveld and co- workers
(Zaneveld and Bartz, 1984; Zaneveld et al., 1990). One of the greatest problems of making
a practical reflective tube is the reflectivity coefficient of the walls. A Monte Carlo study
of the performance of a reflective tube absorption meter (Kir k, 1992) shows that the results
quickly deteriorate with reflectivity decreasing from 100% . As it is virtually impossible to
produce perfect reflecting walls, especially ones that would not deteriorate with prolonged use
of the instrument, the concept of a quartz glass tube surroun ded by air was proposed instead
(Zaneveld et al., 1990). Assuming smooth surfaces of the tub e, all photons encountering the
wall at an angle to the wall surface smaller than the critical angle, 41◦, must be internally
reflected. Therefore, a clean quartz reflection tube should c ollect all photons scattered in the
angular range between 0◦ and 41◦ (if multiple scattering is neglected). However, the large
losses of photons scattered at angles above 41◦ are the main theoretical source of error for
quartz tube absorption meters.
The Monte Carlo calculations by Kirk (1992) were conducted f or a prototype absorption
tube with an almost parallel light beam resembling the Wetla bs ac-9 absorption meter. The
2
PDF page 3
results showed that the relative error of absorption is alwa ys positive and increases linearly
with the ratio of scattering to absorption. The error increa ses with decreasing wall reflectance
and acceptance angle of the receiver. Another study by Hakvo ort and Wouts (1994) used a
Lambertian light source. In this case the absorption error d ecreases with the decreasing angle
of photon acceptance. However, early prototypes of HiStar, the new Wetlabs spectrophotome-
ter, has a diverging beam limited to 20◦. The receiving end of the tube is illustrated in Fig. 1.
It consists of a Light Shaping Diffuser (LSD) in front of a len s. The fiber transmits light to a
spectrometer. Such an arrangement results in a large receiv ing area which through the use of
the LSD and the lens translates into the small acceptance angle needed by the spectrometer.
In this paper we take account of the reflections at both ends ofthe tube, that were neglected
by previous studies of this kind. We also extend previous results by considering a more realistic
arrangement, by introducing weighting functions that quan titatively show the scattering error
as a function of angle, and by providing calculations for some cases of practical interest.
2 CALCULATION SETUP
The Monte-Carlo code used was adopted from the code written to determine the effects of self-
shading on an in-water upwelling irradiance meter (Piskozub 1994). This is a forward Monte
Carlo algorithm, meaning that the photons are traced in the forward direction starting from the
light source. An absorption event ends a photon’s history. The low values of the optical depth
inside the absorption tube make this approach comparatively efficient.
The absorption tube studied (called henceforth α-TUBE) is a cylindrical shell of inner
radius r=0.006 m and length d=0.23 m which corresponds to the dimensions of the HiStar
quartz reflective tube (see Fig. 1). Thickness of the quartz wall is ∆r=0.002 m. The indices of
refraction used are 1.33 for water and 1.41 for quartz.
The photons inside the α-TUBE are traced along their three dimensional trajectories. The
scattering or absorption events inside the volume of the wat er sample are defined by inherent
optical properties of the medium: the absorption coefficien t of the medium a, scattering coef-
ficient b, and the scattering phase function β(θ). Random numbers are used to choose whether
a given photon traveling from point (x,y,z) towards direction (θ, ϕ) will reach the wall or the
end of the cylinder. Otherwise, it is assumed that the photon ends its trajectory inside the liq-
uid medium, and the type of the event (scattering or absorption) is determined by comparing a
random number to the single scattering albedo ω0 value. All random number used in the code
are uniformly distributed in the open interval (0, 1).
We define the Cartesian coordinate system such that the axis of the tube is co-located with
the z-axis, and the center of the source end of the tube defines the origin of the coordinate
system. Therefore, θ is the angle between the photon direction and the tube axis an d ϕ is the
angle of projection of the photon direction onto the x-y plan e. In every scattering event the
new direction of the photon is chosen using the relevant phase function.
The source of photons is assumed to be within a circle of the ra dius r = 1.5 mm emitting
photons into a cone of 25◦ half-width. This represents the fiber head. The angular distribution
of photons is assumed to be Lambertian up to the 25◦. The position of the photon entering
the tube itself is calculated, taking into account the dista nce from the fiber head to the tube
entrance (0.019 m) and refraction from air to water at the mouth of the reflective tube.
3
PDF page 4
Figure 1: Schematic representation of the reflective tube ab sorption meter setup discussed in the
paper
Such an α-TUBE, defined above, is similar to the prototype WET Labs HiStar, but it does
not precisely model the detailed radiance structure surrounding both the source and the receiver
ends of the reflective tube. We believe that those details areof minor importance in comparison
to neglecting the albedo of the tube ends. Therefore we intro duced the albedo of the source
end for photons returning to it from inside the tube, for whic h we estimated the albedo as 0.3
(0.2 being the diffuse albedo and 0.1 the specular albedo). T his is a rough assumption of the
combined effects of the input window and the silver colored m etal surface around the fiber
head.
The quartz walls are treated as ideally smooth, reflecting and refracting photons according
to geometrical optics. The direction of a refracted photon is determined in the code by Snell’s
law and probability of reflection and refraction by Fresnel’ s law. However, no polarization
effects are included. The photons are also followed inside t he quartz, which is assumed to be
non-absorbing. All photons leaving the outer surface of the quartz tube are treated as lost. The
tube is assumed to be surrounded by an ideally black medium.
The receiver end of the tube is treated as a smooth surface refl ecting 8% of the incident
photons. This corresponds to the angular average of reflecti on for the glass window and the
Light Shaping Diffuser (LSD) used. It is assumed that a photon hitting the diffuser plate at an
angle θ reaches the receiver with a probability proportional to theangular diffusion distribution
(or scattering angle profile) of the LSD. The Light Shaping Diffuser used in the instrument is a
Physical Optics Corporation 20◦ LSD. Also, we use in the calculations two other diffusers:60◦
LSD and 95◦ LSDs. The angles are half-widths of the angular diffusion fu nction. This means
that, for example, a 60◦ LSD has a 50% transmission efficiency at angle 30◦. The angular
diffusion distribution used was determined by the least-sq uare fits of the experimental curves
provided by the producer. The 20◦ LSD was approximated by two Gaussian functions (with
all angles in degrees):
L20(θ) = A exp
[
−0.5
( θ
B
) 2]
+ (1 − A)
[
−0.5
( θ
C
) 2]
(1)
where A = 0.85, B = 8.0◦, C = 15.3◦. The 60◦ LSD was approximated by a single Gaussian
4
PDF page 5
Figure 2: Light diffusing characterisitics of the Light Shaping Diffusers for three different types of
diffusers.
function:
L60(θ) = A exp
[
−0.5
( θ
B
) 2]
(2)
where A = 1 .0 and B = 24 .5◦. Finally, the 95◦ LSD was approximated by a Gompertz
function (Gompertz, 1825):
L95(θ) = A exp
[
− exp
(
−
( θ − θ0
B
))]
(3)
where A = 1.0, B = 7.33◦ and θ0 = 50.75.
The shapes of the three LSD diffusion angle profiles are presented in Figure 2. The standard
Petzold scattering phase function for open turbid waters wa s used (Petzold 1972) except for
the calculations of the effect of the phase function shape on the measurement error where
Henyey-Greenstein phase functions (Henyey and Greenstein, 1941) were used.
Each Monte Carlo calculation was performed twice: one to det ermine how many photons
are recorded if there is no absorption and scattering ( P0) and the other for the IOPs (absorp-
tion, scattering and phase function) of the aquatic medium b eing studied (P). We define the
5
PDF page 6
measured absorption am as Kirk (1992):
am = (1/d) ln(P0/P ) (4)
where d is the length of the cylinder.
It must be noted that the first program run corresponds to cali bration of the instrument in
“ideal water,” not in air. We use the water index of refractio n for this medium so as not to
influence the new photon direction in refraction events. The refore, the results are different
from what one would obtain if the tube was filled with air. Such a non-scattering and non-
absorbing liquid medium does not exist. However, it is the mo st obvious reference value to
use in the calculations aimed at studying the calibration of the instrument, because it does
not introduce any photon losses due to scattering (b=0) and d oes not change the behavior of
photons on the quartz-liquid border.
3 RESULTS
3.1 SEMI-ANALYTICAL CONSIDERATIONS
In this section we present qualitative arguments related to the optical phenomena influencing
the fate of photons in the absorption tube. The purpose of this is to provide approximate results
that can be compared with exact Monte Carlo results later. Th e semi- analytic results give a
better understanding of the underlying physics. If, for exa mple, the IOPs of the studied water
sample are assumed to be: b=0.8 m − 1, a=0.2 m− 1, and the Petzold San Diego Harbor (turbid
water) phase function, one can expect that 4.5% ( exp(−ad)) of all photons will be absorbed
and 16.8% (exp(−bd)) will be scattered in the water volume inside the quartz tube.
The phase function used, determines that 92.9% of all scatte ring events take place in the
0-41◦ range. Assuming for simplicity, that before scattering, al l photons traveled parallel to
the tube axis, all those scattered photons will be reflected b ack into the tube on all encounters
with the tube wall. Similarly, all photons backscattered in to the range of 139-180◦ stay in the
tube on their way back to the source end. These represent 0.3% of all photons. The remaining
6.8% of all scattered photons are scattered into the range 41-139◦. These photons have a large
probability of leaving the tube on the first encounter with th e wall. The angles to the z-axis
at which those photons travel make it virtually certain that they will leave the tube before
reaching either end, unless they are scattered close to 41◦. For example at θ = 45 ◦, a photon
scattered in the very center of the tube will hit the wall 9 tim es before reaching the end of the
tube making the probability of not leaving it < 10− 8. The 6.8% of scattered photons that are
lost translates into 1.1% of all traced photons being lost through the walls.
After the absorption and scattering losses are taken into ac count, 94.3% of all traced pho-
tons reach the receiver end of the tube. Thus, the assumption of 8% specular albedo of the
receiver end of the tube means that 7.5% of all traced photons are reflected from the receiver
end of the tube. Absorption and scattering losses result in o nly 7.1% of all initial photons
coming back to the source end after being reflected. The backs cattered photons, which are
attenuated in a pathlength approximately equal to the tube l ength, increase this number to
7.2%.
6
PDF page 7
Another important aspect of a quartz absorption tube is the length of the path that the pho-
tons travel in quartz instead of water. Any distance traveled in quartz decreases the attenuation
due to the negligible absorption coefficient in the visible r ange of the spectrum (Kirk 1992).
This decreases the measured absorption values because the r eceiver is reached by more pho-
tons due to the shorter path through the absorbing medium. In the studied tube geometry, the
ratio of tube diameter to wall thickness is only 6:1, and considering that every time the photon
goes across the water volume it needs to cross the wall twice ( out and back into the tube) it
would seem that this effect must be overwhelming. However, d ue to Snell’s law and the nar-
row angles between the direction of most photons and the wall, the average path of photons in
quartz is much smaller.
3.2 SANITY CHECKS OF THE MONTE CARLO RESULTS
Even if rough approximations of the results are possible wit h simple arithmetic calculations,
Monte Carlo modeling includes all optical phenomena taking place inside the tube. This can
be illustrated by comparing the approximate values from the previous section to the results of
a Monte Carlo run using 40 million photons with the same IOPs. The number of photons lost
through the walls is 1.74% (estimated above as 1.1%). The dis crepancy can be explained by
the divergence of the input beam: photons entering the tube a t an angle may leave it through
the walls, even if scattered from its original direction at l ess than the angle of total internal
reflection, resulting in higher scattering losses.
The Monte Carlo derived absorption losses percentage is 4.0 9% (semi-analytically esti-
mated as 4.5%). The reason for this is that the path length of t he photons is in quartz instead
of in water. The Monte Carlo code makes it possible, by tracin g each photon individually,
to calculate the average path traveled by the photons in wate r to be 21.15 cm and in quartz:
2.83cm. This means that the path in water is 8% shorter than th e length of the reflective tube,
a value much higher than that calculated by Kirk (1992) for a parallel beam in which only the
scattered photons encounter the walls. The losses by absorp tion on the front end of the tube
are 4.84% of all photons (estimated as 5.0%). The small difference is the net result of smaller
absorption losses and greater scattering losses.
The effect of the shortened optical path, described above, i s not very grave if the way the
instrument is calibrated is taken into account. All calibration techniques involve a comparison
of the received signal (that is the number of photons reachin g the receiver) for the measured
sample with pure water or air values. Using pure water almost completely removes the error
due to non-scattered photons traveling partly inside the wall, as the effect will be identical for
water of any IOPs. Calibrating the instrument in air leaves s ome error due to path length in
quartz because air has a different index of refraction making the path length in quartz shorter.
There is, however, a much bigger source of error in calibrati ng the instrument without water,
which is the removal of total internal reflection on the quartz wall by surrounding both its sides
by the same medium.
3.3 V ARIATION IN THE SCATTERING-ABSORPTION RATIO
It was shown by Kirk (1992) that for a reflective tube absorption meter propagating an almost
parallel beam, that the ratio of the measured to the true valu e of the absorption coefficient
7
PDF page 8
am/a increases linearly with b/a at a rate depending on the phase function used. We decided
to test whether such a relationship will hold for the relatively divergent light beam and limited
acceptance angle of the α-TUBE. Figure 3 shows that this is indeed the case for all stud ied
angles of acceptance (depending on the LSD type used). The slope coefficients w of the linear
fit
am = a + wb (5)
are 0.194 for 20◦ LSD, 0.139 for 60◦ LSD, and 0.0957 for 95◦ LSD. These results suggest that
the scattering losses of the instrument increase with the de creasing angle of acceptance of the
receiver. It must be noted that the values of the coefficient w depend on the scattering phase
function used.
Figure 3 shows that there is a small offset in Equation 5. Its general version am = a+wb+o
(where o is the offset) can be transformed to
a = am − wc − o
1 − w (6)
which, for a given phase function and absorbing tube geometr y, allows for determining the
true a value if attenuation c is measured independently. The parameters w and o may be
calculated for the given absorption meter geometry if the ph ase function is known at least
approximately. The values of the coefficient w, for the physically important range of Henyey-
Greenstein asymmetry parameter g and the 20◦ LSD receiver are shown in Figure 4.
3.4 V ARIATION IN THE SCATTERING PHASE FUNCTION
The measurement error in absorption depends on the scatteri ng phase function of the aquatic
medium (Kirk, 1992). To study the effect of the scattering ph ase function on absorption mea-
sured by the given instrument setup with 20◦ LSD, we performed a sequence of Monte Carlo
calculations for the Henyey-Greenstein phase function (Henyey and Greenstein 1941) by vary-
ing the asymmetry parameter g
β(θ) = 1 − g2
(1 + g2 − 2g cos θ)3/2 (7)
where g is a parameter determining the shape of the phase func tion. In this paper we assume
range of g from g = 0 (isotropic scattering) to g = 1 (forward scattering).
The results in Fig. 5 show am/a as a function of the asymmetry parameter. They were
obtained by running the code for the studied tube equipped wi th three different LSD plates
(20◦, 60◦, and 95◦). As expected, for g approaching 1 the am/a ratio is close to 1 as all
photons are scattered almost in the direction that they were headed before the scattering. For a
more realistic range of g values between 0.8 and 0.9 (Mobley, 1994) the measured absorption
increases when g decreases due to the greater scattering los ses of the photons. The value of
am/a < 1 for g = 0 .95 and 95◦ LSD is not a statistical error, but the effect of the slightly
longer path of scattered photons in the quartz, in comparison to the non-scattering ”ideal water”
which results in decrease of absorption.
For small g values the am/a values (Fig. 5) approach the maximum (a+b)/a at which all
the scattered photons are lost and the measured value of absorption equals the total attenuation.
8
PDF page 9
Figure 3: Ratio of the measured to the true absorption coeffic ient am/a as a function of the
scattering-absorption ratio b/a and a = 0 .2m− 1, Petzold ”turbid” phase function for three dif-
ferent Light Shaping Diffusers in front of the receiver lens.
9
PDF page 10
Figure 4: The slope coefficient w as a function of Henyey-Gree nstein asymmetry factor g for the
20◦ LSD receiver geometry.
10
PDF page 11
Figure 5: Ratio of the measured to the true absorption coefficient am/a as a function of the Henyey-
Greenstein scattering phase function asymmetry parameter g; a = 0 .2m− 1, b = 0 .8m− 1, Petzold
turbid phase function, for three different Light Shaping Diffusers in front of the receiver lens. Solid
lines are third order polynomial fit.
11
PDF page 12
Figure 6: Ratio of the measured to the true absorption coefficient am/a as a function of the accep-
tance angle of the receiver, a = 0.2m− 1, b = 0.8m− 1, Petzold turbid phase function.
In the case studied (a+b)/a = 5. However, even for an isotropic scattering phase function (g=0)
the value of the scattering error does not reach this maximum value because some photons are
scattered at angles small enough to be recorded by the receiv er. Again, the figure shows that
increasing the angle of acceptance of the receiver decreases the absorption error for all realistic
phase functions.
3.5 V ARIATION IN THE ACCEPTANCE ANGLE
Results presented by Kirk (1992) for a reflective tube absorption meter with an almost parallel
beam show a rapidly increasing value of am/a with the acceptance angle decreasing below
90◦. On the other hand, Hakvoort and Wouts (1994) suggest that fo r the reflective tube with
Lambertian light source and metallic reflective walls, the m easurement error increases with
the acceptance angle of the receiver. The Monte Carlo code wa s run for several values of the
receiver angle of acceptance to test if the similiar situati on arises for the α-TUBE. No Light
Shaping Diffuser was used in this case. The results are presented in Figure 6.
It can be seen that for large acceptance angles (> 40◦) the am/a values have a similar shape
12
PDF page 13
to that of a parallel beam case. However, for acceptance angles in the 0−40◦ range a minimum
can be observed, instead of a high peak as in the case of a paral lel beam. The reason for that
is that the source beam is more divergent than the angle of acc eptance. For small acceptance
angles the receiver accepts some scattered photons that would not be accepted if they reached
the receiver without a scattering event. This is because the ir angle of incidence is reduced by
the scattering event. This reduces the scattering error of t he absorption measurement. Unlike
the Lambertian source case, there is no major increase of am/a values for large acceptance
angles (although a small increase above 120◦ may be discerned). It should be noted that the
LSD plate used in the instrument corresponds roughly to an acceptance angle of 20◦. Although
this angle is close to a local minimum of am/a, it is still within the range of the highest levels
of absorption error.
3.6 ANGULAR FUNCTION OF PHOTON LOSS PROBABIL-
ITY W (θ)
Full understanding of the mechanism of scattering loss is not possible without determining the
angular relationship of the scattering error of the reflective tube absorption meter. The fraction
of scattered light lost because of the absorption after bein g scattered at a given angle can be
defined as a loss function W (θ). The angular integral of this function multiplied by the pha se
function is the error of the measured absorption value ε = am − a
ε =
∫ π
0
W (θ)β(θ) sin(θ)dθ (8)
The advantage of the W function is that it is not dependent on t he shape of the phase
function but only on the tube geometry. This is true only if mu ltiple scattering is neglected
because a second scattering event influences the fate of a pho ton already scattered. In order
to differentiate among the possible sources of the scatteri ng loss error of the absorption value
we decided to calculate three variants of the W function (a) W0(θ) for scattering losses due to
photons lost through the tube walls, (b) W1(θ) same as W0(θ) and for scattering losses due to
additional photons absorbed by the source (”front”) end of the tube, (c) W2(θ) same as W1(θ)
and for scattering losses due to additional photons lost on the LSD.
By ”additional photons” we mean that only the difference between the number of photons
lost by the absorption and scattering and the ”ideal” water i s taken into account. In each
of the three cases the actual value of W for angle range is deri ved by dividing the number of
photons scattered into the angular sector and subsequentlylost by the sum of these photons plus
the number of photons scattered into the same sector which ar e subsequently recorded by the
receiver. Fig. 7 presents the W functions calculated by running the code for 40 million photons
with a = 0 .2m− 1, b = 0 .8m− 1, Petzold turbid phase function, and 20◦ LSD. In this case the
number of multiple scattered photons is 7% of all which are sc attered. That means that the
influence of the phase function shape on the W function is of secondary importance. All three
functions have the sigmoid shape for the forward scattered p hotons between 0 − 90◦. There
is no sharp step at 41◦ and 139◦ because the divergence of the beam causes some photons to
travel at an angle to the tube axis before scattering which allows photons to escape through the
wall even if scattered at an angle smaller than the total inte rnal reflection angle. On the other
hand, a photon scattered at more than 41◦ back towards the axis may survive the encounter
13
PDF page 14
with the wall without being refracted out of the tube. The sym metrical shape of W0 is caused
by the fact that losses on the ”front” end are not taken into ac count. A photon backscattered
at angle 180 − φ has the same chance of being lost through the walls as the one s cattered at
φ. The symmetry is broken only by the assumed diffusive albedoof the light source end of the
tube because some photons leave the tube by the wall after being diffused on its source end.
Adding the photon losses on the light source end of the tube (f unction W1) changes the
symmetry. Most of the backscattered photons are lost in this way as well as some of the
forward scattered. The latter is caused by two reasons. Some forward-scattered photons are
reflected by the receiver end of the tube back towards the sour ce end. However, it is even
more important that some of the forward scattered photons ar e absorbed on the ”front” end
because they were scattered after they were reflected on the receiver end. The reverse direction
of those photons before scattering complicates the W (θ) function, but we decided to include
these scattering events in the statistics used to calculate the functions because they contribute
to the total scattering losses.
Receiver end losses are caused by more photons missing the re ceiver due to reaching it at
a wider angle after a scattering event. This changes the pict ure by, paradoxically, decreasing
the photon losses for photons that were scattered in the forward direction (function W2). This
means that scattering a photon by a small angle increases its probability of being accepted by
the receiver compared to the average of non-scattered photo ns. This is caused again by the
source beam being wider than the acceptance angle of the rece iver. The discrepancy between
W1 and W2, especially at scattering angles close to 0, is also partly caused by the inclusion of
photons scattered on their way back to the source end of the tu be. One of its consequences is
that the population of photons scattered at close to 0 degree is different from the population of
non-scattered photons.
The shape of the W functions for angles greater than90◦ is not important for the scattering
error estimates because only a fraction of all the photons isbackscattered. For the W2 function
which includes all the photon losses due to scattering, the l oss of backscattered photons is
close to 100%. Therefore, the best fit approximations to W0, W1 and W2 were calculated for
the range 0 − 90◦ only. The three functions were fitted with a 4-parameter sigmoid
W (x) = y0 + A
(1 + exp(−(x − x0)/B)) (9)
where for W0 we have A = 0.935, B = 2 .53◦, x0 = 26 .72◦, y0 = 0 .061, for W1 we have
A = 0.841, B = 3.35◦, x0 = 25.75◦, y0 = 0.157, and for W2 we have A = 0.901, B = 2.58◦,
x0 = 26.79◦, y0 = 0.095.
4 CONCLUSIONS
Monte Carlo calculations of the scattering error for the HiStar prototype absorption meter show
that the different blueprint of the instrument (more divergent light beam and a limited angular
view of the receiver) in comparison to other reflective tube a bsorption meters does influence
the error. Due to additional scattering losses by the view li mited receiver the absorption error
is greater for all combinations of moderate optical parameters. It is possible to correct the error
by careful calibration, and the results presented in this paper for the turbid water Petzold phase
14
PDF page 15
Figure 7: The photon loss probability W (θ) as a function of the scattering angle, a = 0 .2m− 1,
b = 0 .8m− 1, Petzold turbid phase function. These three functions repr esent (a) W0 which defines
losses of photons leaving the tube through the quartz walls, (b) W1 same as W0 and losses on the
light source end of the cylinder, (c)W2 which is the same asW1 but with losses on the receiver-end.
15
PDF page 16
function are a step in this direction. Independent measurements of attenuation may improve the
error correction for water samples of known (at least approximately) phase functions as shown
in the paper. However, the source of inter-instrumental discrepancy in absorption measurement
that is more difficult to correct is their different response s to phase function variability. One
possible solution is to use the photon loss function W. If the phase function of the studied
sea-water is known (or is possible to be estimated by compari ng it to known phase functions
of similar water samples), the W function makes it possible t o calculate directly the value of
the scattering error for the sample.
Our calculations show that the main source of the absorption error is the view limited
receiver. Therefore, we suggest that if technically possib le (it would diminish the amount of
light collected by the receiver optical fiber) a diffuser of w ider angular characteristics should
be used in the instrument receiver setup.
5 REFERENCES
Bricaud A., Babin M., Morel. A., and Claustre H., 1995: ”Vari ability in the chlorophyll-
specific absorption coefficients of natural phytoplankton: analysis and parameterization,” J.
Geophys. Res. , 100, 13321–13332.
Gompertz B., 1925: ”On the Nature of the Function Expressive of the Law of Human Mortal-
ity,” Phil. Trans. Roy. Soc. London., 115, 513.
Gordon H. R. and Morel, A. 1983: ”Remote assessment of ocean c olor for interpretation of
satellite visible imagery, a review” in Lecture notes on coa stal and estuarine studies, vol. 4,
Springler Verlag, New York, 114pp.
Hakvoort J.H.M, Wouts R. 1994: ”Monte Carlo modelling of the light field in reflective tube
type absorption meter,” Proc. SPIE , 2258, Ocean Optics XII, 529–538.
Henyey, L. G., and J. L. Greenstein, 1941: ”Diffuse radiation in the galaxy,” Astrophys. J., 93,
70–83.
Kirk, J.T.O, 1992: ”Monte Carlo modeling of the performance of a reflective tube absorption
meter,” Appl. Opt., 31, 6463–6468
Mobley, C. D., 1994, Light and water: radiative transfer in n atural waters, Academic Press,
San Diego.
Petzold, T. J., 1972: ”V olume scattering functions for selected ocean waters,” SIO Ref. 72-78,
Scripps Institution of Oceanography, Univ. of California, San Diego.
Piskozub J., 1994: ”Effects of surface waves and sea-bottom on self-shading of in-water opti-
cal instruments” in Ocean Optics XII, Proc. SPIE, 2258, 300–308.
16
PDF page 17
Pope R.M., Fry E.S., 1997: ”Absorption spectrum (380-700 nm) of pure water. II. Integrating
cavity measurements,” Appl. Opt., 36, 8710–8723.
Zaneveld J.R.V , Bartz R, 1984: ”Beam attenuation and absorp tion meters” in Ocean Optics
VII, M.A. Blizard ed., Proc. SPIE, 489, 318–324.
Zaneveld J.R.V , Bartz R., Kitchen J.C, 1990: ”A reflective tu be absorption meter,” in Ocean
Optics X, R.W. Spinard ed., Proc. SPIE, 1302, 124-136.
17